Table 1. Performance of the Panbio COVID-19 Ag Rapid Test Device for SARS-CoV-2 detection

Testing method Real-time RT-PCR Sensitivity (95% CI) Specificity (95% CI)
P N
Panbio COVID-19 Ag P 15 0 50.0% (31.3−68.7) 100.0% (88.3−100.0)
N 15 30
P, positive; N, negative.